AFM University Introduction to Atomic Force Microscopy by Paul West

« Cover
« Foreward
« Chapter 1
 « Chapter 2
 
2.1. Basic Concepts
2.2. The AFM Stage
2.2.1 XYZ Scanners
2.2.2 Force Sensors
2.2.3 Integrating LL-Force Sensors and Scanners
2.2.4 Z Motors-Probe Approach
2.2.5 X-Y Stage
2.2.6 Optic Microscope
2.2.7 Mechanical Loop
2.3 Electronics
2.4 AFM Acquisition Software
2.5 LL-AFM Cantilevers and Probes
2.6.1 Vibrations
2.6.2. Environmental Scanning
2.6.3 Heating/Cooling Stages
2.6.4 Higher Speed AFM Scanning
 « Chapter 3
 « Chapter 4
 « Chapter 5
 « Chapter 6
 « Chapter 7
« Appendix A
« Appendix B
« Appendix C
« Appendix D
About
Downloads
Home


Chapter 2


next » index « back

b) Mechanical Vibrations are transmitted to the AFM stage from mechanical contact. Sources of mechanical noise include floor vibrations. Usually floor vibrations are greater the higher up in a building the AFM is located. Mechanical vibrations are reduced by putting the AFM on a vibration table.
FIGURE 2-44Left: Vibration tables have a transmission curve that shows the attenuation of the table to vibrations as a function of frequency. There is always a natural resonance associated with the vibration table that is typically below 5 Hz. Right: Acoustic vibrations are reduced using a vibration enclosure.
2.6.2. Environmental Scanning
Scanning in a controlled environment with an AFM can be advantageous for many reasons:
  • Reactive materials are stable in an inert environment.
  • Biological materials can be measured in situ.
FIGURE 2-45 Left: A microcell is created by gluing the cantilever/substrate on to a piece of glass. Right: An environmental cell is created by making a sealed cell that completely surrounds the probe/cantilever and the sample.
There are many ways for controlling the environment of scanning with an AFM. The first is to place the AFM in a vacuum chamber and remove all of the gasses from the microscope stage. The second is to put the AFM in a controlled environment such as a glove box. The third is to put the sample and AFM tip in a liquid environment. Figure 2-45 shows two methods for putting the sample/probe in a liquid environment.
next » « back
  58