AFM University Introduction to Atomic Force Microscopy by Paul West

« Cover
« Foreward
« Chapter 1
 
1.1. History
1.2. Comparison
1.3. Enabling Nanotechnology
 « Chapter 2
 « Chapter 3
 « Chapter 4
 « Chapter 5
 « Chapter 6
 « Chapter 7
« Appendix A
« Appendix B
« Appendix C
« Appendix D
About
Downloads
Home


Chapter 1


next » index « back

SEM/TEM instruments are capable of doing much more than topography measurements. For example, electron beam instrumentation can do EDX measurements or even electron beam initiated lithography. Likewise, the AFM can make many types of measurements other than AFM topographical measurements. For example, AFM instruments can make thermal, magnetic and electric fields maps of a surface. Like the SEM/TEM, an AFM can also initiate lithographic changes on a surface.
Although the time required for making a measurement with the SEM image is typically less than an AFM, the amount of time required to get meaningful images is similar. This is because the SEM/TEM often requires substantial time to prepare a sample. With the AFM little and often no sample preparation is required. Figure 1-9 shows the comparison between a TEM image of nano-particles and the AFM image of the same nanoparticles.
FIGURE 1-9 AFM image, color, and TEM, grayscale, of 500 nm diameter particles. A line profile from the AFM image shows the height of the particles.
FIGURE 1-10 Comparison of the time for measurements and instrumentation cost of optical, AFM, and SEM/TEM microscopes.
In comparison with an optical microscope and the SEM/TEM an AFM is more difficult to use than the optical microscope and easier to use than the SEM/TEM. Figure 1-10 compares the relative time and cost for optical, AFM, and SEM/TEM microscopes.
Lastly, an optical microscope requires the least amount of laboratory space, while the SEM/TEM requires the most amount of laboratory space. An AFM is in the middle of these two.
next » « back
  9